Multipurpose tool for QC testing of radiation fields used for stereotaxy. It allows to measure all standard quality assurance parameters very quickly and comfortably.
MEASURED AND EVALUATED PARAMETERS :
- Absolute dose at CAX
- Beam energy stability
- Relative dose analyzes - symmetry, flatness
- Light and radiation fields coincidence
DETECTORs :
- 64 Diode Detectors (DD) - active area 0.5 x 0.5 mm
- Beams: photons 1-25MV / electrons 4-25MeV
- Dose Rate Range: 0.01÷30 Gy/min (resol. 0.1%)
DESIGN :
- GRID pattern with spacing 1.5 mm
- Tested Field Sizes: Circular - max. 2 cm
- Dimensions / Weight: 40x130x6 mm / 0.1kg
UNBEATABLE POWER / PRICE RATIO
|
|
4000 - DETECTORs - MATRIX Systems (Series 4000)
4500 - MATRIX Systems for PENCIL Beams Mapping
COMING SOON
MR 4521 DD - 2D-MATRIX – STAR pattern – 64 DIODEs
» for QC Tests of Very Small Radiation Fields (Pencil beams)
» for QC Tests of Very Small Radiation Fields (Pencil beams)
COMING SOON
MR 4531 IC - 2D-MATRIX – GRID pattern – 64 ION.CHAMBERs
» for QC Tests of Very Small Radiaton Fields (Pencil beams)
» for QC Tests of Very Small Radiaton Fields (Pencil beams)
Multipurpose tool for QC testing of radiation fields used for stereotaxy. It allows to measure all standard quality assurance parameters very quickly and comfortably.
MEASURED AND EVALUATED PARAMETERS :
- Absolute dose at CAX
- Beam energy stability
- Relative dose analyzes - symmetry, flatness
- Light and radiation fields coincidence
DETECTORs :
- 64 Ionization Chambers (IC) - diameter 2 mm
- Beams: photons 1-25MV / electrons 4-25MeV
- Dose Rate Range: 0.1÷30 Gy/min (resol. 0.1%)
DESIGN :
- GRID pattern with spacing 3 mm
- Tested Field Sizes: Circular - max. 2 cm
- Integrated Barometer and Thermometer
- Dimensions / Weight: 40x130x6 mm / 0.1kg
UNBEATABLE POWER / PRICE RATIO
|
|