4000 - DETECTORs - MATRIX Systems (Series 4000)
4300 - MATRIX Systems for MEDIUM Beams Mapping
show product details hide product details PRELIMINARY    MR 4321 DD - 2D-MATRIX – STAR pattern – 241 DIODEs
» for Daily QC Checks of Beams used in Radiotherapy
2D-MATRIX – STAR pattern – 241 DIODEs
Multipurpose tool for QC testing of radiation fields. It allows to measure all standard quality assurance parameters very quickly and comfortably.
 
MEASURED AND EVALUATED PARAMETERS :
- Absolute dose at CAX
- Beam energy stability
- Relative dose analyzes - symmetry, flatness
- Light and radiation fields coincidence
 
DETECTORs :
- 241 Diode Detectors (DD) - active area  0.5 x 0.5 mm
-   13 Ionization Chambers (IC)
- Beams:  photons 1-25MV  /  electrons 4-25MeV
- Dose Rate Range:  0.01÷30 Gy/min  (resol. 0.1%)
 
DESIGN :
- STAR pattern:
     61 DDs in each profile with spacing 2.5 / 5 mm
       7 DDs in each field penumbra (w. spacing 2.5)
    12 ICs for beam energy stability check
      1 IC at CAX for beam output stability check
- Tested Field Sizes: 5x5 / 10x10 / 15x15 /20x20 cm
- Integrated Barometer and Thermometer
- Dimensions / Weight:  250x300x15mm  / 1.5kg
 
 
UNBEATABLE  POWER / PRICE  RATIO
show product details hide product details     2018 AUGUST MR 4321 IC - 2D-MATRIX – STAR pattern – 512 ION.CHAMBERs
» for Daily QC Checks of Beams used in Radiotherapy
 2D-MATRIX – STAR pattern – 512 ION.CHAMBERs
Multipurpose tool for QC testing of radiation fields. It allows to measure all standard quality assurance parameters very quickly and comfortably.
 
MEASURED AND EVALUATED PARAMETERS :
- Absolute dose at CAX
- Beam energy stability
- Relative dose analyzes - symmetry, flatness
 
DETECTORs :
- 512 Ionization Chambers (IC) - rounded 2x4 mm
-   13 Ionization Chambers (IC) - diameter 3 mm
- Beams:  photons 1-25MV  /  electrons 4-25MeV
- Dose Rate Range:  0.1÷30 Gy/min  (resol. 0.1%)
 
DESIGN :
- STAR pattern:
     91 ICs in each CAX profile with spacing 2.5
     21 ICs in each Off-CAX profile with spacing 2.5 / 5 mm
       7 ICs in each field penumbra (w. spacing 2.5)
    12 ICs for beam energy stability check
      1 IC at CAX for beam output stability check
- Tested Field Sizes: 5x5 / 10x10 / 15x15 /20x20 cm
- Integrated Barometer and Thermometer
- Dimensions / Weight:  300x500x15mm  / 1.5kg
 
 
UNBEATABLE  POWER / PRICE  RATIO