4000 - DETECTORs - MATRIX Systems (Series 4000)
4200 - MATRIX Systems for LARGE Beams Mapping
show product details hide product details PRELIMINARY  MR 4201 DD - 1D-MATRIX – LINEAR pattern – 220 DIODEs
» for High-resolution measurements of Beam Profiles
1D-MATRIX – LINEAR pattern – 220 DIODEs
Avanced tool for QC testing of radiation fields. It allows to measure profile both for open and for wedged fields very quickly and comfortably.
 
MEASURED AND EVALUATED PARAMETERS :
- Beam profile
- Light and radiation profiles coincidence
 
DETECTORs :
- 220 Diode Detectors (DD) - active area  0.5 x 0.5 mm
   Detector spacing is 2 mm
   Max. profile lenght 44 cm
- Beams:  photons 1-25MV  /  electrons 4-25MeV
 
DESIGN :
- LINEAR pattern
- Dimensions / Weight:  450x15x15mm  / 0.5kg
 
UNBEATABLE  POWER / PRICE  RATIO
show product details hide product details PRELIMINARY  MR 4201 IC - 1D-MATRIX – LINEAR pattern – 176 ION.CHAMBERs
» for High-resolution measurements of Beam Profiles
1D-MATRIX – LINEAR pattern – 176 ION.CHAMBERs
Avanced tool for QC testing of radiation fields. It allows to measure profile both for open and for wedged fields very quickly and comfortably.
 
MEASURED AND EVALUATED PARAMETERS :
- Beam profile
 
DETECTORs :
- 176 Ionization Chambers (IC) - active area  2 x 4 mm
   Detector spacing is 2.5 mm
   Max. profile lenght 44 cm
- Beams:  photons 1-25MV  /  electrons 4-25MeV
 
DESIGN :
- LINEAR pattern
- Dimensions / Weight:  450x15x15mm  / 0.5kg
 
UNBEATABLE  POWER / PRICE  RATIO
show product details hide product details COMING SOON      MR 4234 IC - 2D-MATRIX – GRID pattern – 2300 ION.CHAMBERs
» for IMRT verification and QC testing of Beams used in Radiotherapy
2D-MATRIX – GRID pattern – 2300 ION.CHAMBERs
Multipurpose tool for IMRT verification and for QC testing of large / medium size radiation fields used for radiotherapy. It allows to measure all standard quality assurance parameters very quickly and comfortably.
 
MEASURED AND EVALUATED PARAMETERS :
- IMRT patient treatment plans verification
- Absolute dose at CAX
- Beam energy stability
- Relative dose analyzes - symmetry, flatness
 
DETECTORs :
- 2298 Ionization Chambers (IC) - diameter 4 mm
- Beams:  photons 1-25MV  /  electrons 4-25MeV
- Dose Rate Range:  0.1÷50 Gy/min  (resol. 0.1%)
 
DESIGN :
- GRID pattern with spacing 10 mm  /  in HD area 7 mm
- Tested Field Sizes:  max. 35x40 cm
- High density (HD) area :  30x34 cm
- Integrated Barometer and Thermometer
- Dimensions / Weight:  370x530x38 mm  / 8kg
 
UNBEATABLE  POWER / PRICE  RATIO
4300 - MATRIX Systems for MEDIUM Beams Mapping
show product details hide product details PRELIMINARY    MR 4321 DD - 2D-MATRIX – STAR pattern – 241 DIODEs
» for Daily QC Checks of Beams used in Radiotherapy
2D-MATRIX – STAR pattern – 241 DIODEs
Multipurpose tool for QC testing of radiation fields. It allows to measure all standard quality assurance parameters very quickly and comfortably.
 
MEASURED AND EVALUATED PARAMETERS :
- Absolute dose at CAX
- Beam energy stability
- Relative dose analyzes - symmetry, flatness
- Light and radiation fields coincidence
 
DETECTORs :
- 241 Diode Detectors (DD) - active area  0.5 x 0.5 mm
-   13 Ionization Chambers (IC)
- Beams:  photons 1-25MV  /  electrons 4-25MeV
- Dose Rate Range:  0.01÷30 Gy/min  (resol. 0.1%)
 
DESIGN :
- STAR pattern:
     61 DDs in each profile with spacing 2.5 / 5 mm
       7 DDs in each field penumbra (w. spacing 2.5)
    12 ICs for beam energy stability check
      1 IC at CAX for beam output stability check
- Tested Field Sizes: 5x5 / 10x10 / 15x15 /20x20 cm
- Integrated Barometer and Thermometer
- Dimensions / Weight:  250x300x15mm  / 1.5kg
 
 
UNBEATABLE  POWER / PRICE  RATIO
show product details hide product details     2018 AUGUST MR 4321 IC - 2D-MATRIX – STAR pattern – 512 ION.CHAMBERs
» for Daily QC Checks of Beams used in Radiotherapy
 2D-MATRIX – STAR pattern – 512 ION.CHAMBERs
Multipurpose tool for QC testing of radiation fields. It allows to measure all standard quality assurance parameters very quickly and comfortably.
 
MEASURED AND EVALUATED PARAMETERS :
- Absolute dose at CAX
- Beam energy stability
- Relative dose analyzes - symmetry, flatness
 
DETECTORs :
- 512 Ionization Chambers (IC) - rounded 2x4 mm
-   13 Ionization Chambers (IC) - diameter 3 mm
- Beams:  photons 1-25MV  /  electrons 4-25MeV
- Dose Rate Range:  0.1÷30 Gy/min  (resol. 0.1%)
 
DESIGN :
- STAR pattern:
     91 ICs in each CAX profile with spacing 2.5
     21 ICs in each Off-CAX profile with spacing 2.5 / 5 mm
       7 ICs in each field penumbra (w. spacing 2.5)
    12 ICs for beam energy stability check
      1 IC at CAX for beam output stability check
- Tested Field Sizes: 5x5 / 10x10 / 15x15 /20x20 cm
- Integrated Barometer and Thermometer
- Dimensions / Weight:  300x500x15mm  / 1.5kg
 
 
UNBEATABLE  POWER / PRICE  RATIO
4400 - MATRIX Systems for SMALL Beams Mapping
show product details hide product details PRELIMINARY      MR 4431 DD - 2D-MATRIX – GRID pattern – 640 DIODEs
» for IMRT verification and QC testing of Beams used in Stereotactic Radiotherapy
2D-MATRIX – GRID pattern – 640 DIODEs
Multipurpose tool for IMRT verification and for QC testing of small radiation fields used for stereotaxy. It allows to measure all standard quality assurance parameters very quickly and comfortably.
 
MEASURED AND EVALUATED PARAMETERS :
- IMRT patient treatment plans verification
- Absolute dose at CAX
- Beam energy stability
- Relative dose analyzes - symmetry, flatness
 
DETECTORs :
- 640 Diode Detectors (DD) - active area  0.5 x 0.5 mm
- Beams:  photons 1-25MV  /  electrons 4-25MeV
- Dose Rate Range:  0.01÷30 Gy/min  (resol. 0.1%)
 
DESIGN :
- GRID pattern with spacing 4 mm
- Tested Field Sizes:  max. 10x10 cm
- Integrated Barometer and Thermometer
- Dimensions / Weight:  150x250x15mm  / 1kg
 
UNBEATABLE  POWER / PRICE  RATIO
show product details hide product details COMING SOON      MR 4431 IC - 2D-MATRIX – GRID pattern – 640 ION.CHAMBERs
» for IMRT verification and QC testing of Beams used in Stereotactic Radiotherapy
2D-MATRIX – GRID pattern – 640 ION.CHAMBERs
Multipurpose tool for IMRT verification and for QC testing of small radiation fields used for stereotaxy. It allows to measure all standard quality assurance parameters very quickly and comfortably.
 
MEASURED AND EVALUATED PARAMETERS :
- IMRT patient treatment plans verification
- Absolute dose at CAX
- Beam energy stability
- Relative dose analyzes - symmetry, flatness
 
DETECTORs :
- 640 Ionization Chambers (IC) - diameter 2 mm
- Beams:  photons 1-25MV  /  electrons 4-25MeV
- Dose Rate Range:  0.1÷30 Gy/min  (resol. 0.1%)
 
DESIGN :
- GRID pattern with spacing 4 mm
- Tested Field Sizes:  max. 10x10 cm
- Integrated Barometer and Thermometer
- Dimensions / Weight:  150x250x15mm  / 1kg
 
UNBEATABLE  POWER / PRICE  RATIO
4500 - MATRIX Systems for PENCIL Beams Mapping
show product details hide product details   COMING SOON MR 4521 DD - 2D-MATRIX – STAR pattern – 64 DIODEs
» for QC Tests of Very Small Radiation Fields (Pencil beams)
2D-MATRIX – STAR pattern – 64 DIODEs
Multipurpose tool for QC testing of radiation fields used for stereotaxy. It allows to measure all standard quality assurance parameters very quickly and comfortably.
 
MEASURED AND EVALUATED PARAMETERS :
- Absolute dose at CAX
- Beam energy stability
- Relative dose analyzes - symmetry, flatness
- Light and radiation fields coincidence
 
DETECTORs :
- 64 Diode Detectors (DD) - active area  0.5 x 0.5 mm
- Beams:  photons 1-25MV  /  electrons 4-25MeV
- Dose Rate Range:  0.01÷30 Gy/min  (resol. 0.1%)
 
DESIGN :
- GRID pattern with spacing 1.5 mm
- Tested Field Sizes:  Circular - max. 2 cm
- Dimensions / Weight:  40x130x6 mm  / 0.1kg
 
 
UNBEATABLE  POWER / PRICE  RATIO
show product details hide product details COMING SOON      MR 4531 IC - 2D-MATRIX – GRID pattern – 64 ION.CHAMBERs
» for QC Tests of Very Small Radiaton Fields (Pencil beams)
2D-MATRIX – GRID pattern – 64 ION.CHAMBERs
Multipurpose tool for QC testing of radiation fields used for stereotaxy. It allows to measure all standard quality assurance parameters very quickly and comfortably.
 
MEASURED AND EVALUATED PARAMETERS :
- Absolute dose at CAX
- Beam energy stability
- Relative dose analyzes - symmetry, flatness
- Light and radiation fields coincidence
 
DETECTORs :
- 64 Ionization Chambers (IC) - diameter 2 mm
- Beams:  photons 1-25MV  /  electrons 4-25MeV
- Dose Rate Range:  0.1÷30 Gy/min  (resol. 0.1%)
 
DESIGN :
- GRID pattern with spacing 3 mm
- Tested Field Sizes:  Circular - max. 2 cm
- Integrated Barometer and Thermometer
- Dimensions / Weight:  40x130x6 mm  / 0.1kg
 
 
UNBEATABLE  POWER / PRICE  RATIO